MassQC Help > Troubleshooting > MS2 Spectra > Triggering          
MS2 Spectra and Triggering

Triggering poses an interesting paradigm with mass spectrometry and, in particular, tandem mass spectrometry (MS/MS). We tune the instrument (and create the method) to trigger on the peaks we want (or the most abundant peaks), but sometimes the triggering occurs on other than desired peaks - on noise, for example.

In this case, we see two metrics out of range, MS1 Total Ion Current and MS2 Peaks Per Spectrum:

MS1 Total Ion Current, MS2 Peaks Per Spectrum

And to confirm the metric's historical data are out of range, we go to the Control Charts to verify:

MS2 Peaks Per Spectrum

So why do we think that these two metrics indicate triggering on noise? Well, we look to the mass spectra, (as is advised when metrics are out of range and it is not obvious what the problem is):

Triggering on Noise

Now, taking a closer look a the noise level, we see that for over 20 minutes, there are not many eluting peptides, yet at the MS2 level, the Peaks per Spectrum is flagged high, as well as the Total Ion Current, both symptoms of higher than normal noise:

High Signal Noise Level

Metrics discussed this section:
Forum

If you have questions, comments and/or insights about MassQC, you can share them on the MassQC Users Forum.